![]() ![]() Non-clinical testing has demonstrated that patients with these specific Grace Medical ossicular implants can undergo MRI safely, immediately after implantation under the following conditions: The MRI Information is, as follows: Imaging at 3 Tesla or Less The Otologic device families listed in Table 2 below are considered MR Conditional according to the American Society for Testing and Materials (ASTM) International, Designation: F2503-13 and based on MRI tests conducted on selected implants that represented worst case with regard to materials and mass. ![]() ![]() The Otologic device families listed in Table 1 below are considered MR Safe according to the American Society for Testing and Materials (ASTM) International, Designation: F2503-13 since these items are made from non-metallic and non conducting materials. Therefore, caution is advised when using MRI procedures in patients with metallic implants and devices. MRI may be contraindicated or can pose risks for patients that have certain metallic implants primarily due to substantial magnetic field interactions or excessive MRI-related heating. Vent Tube MRI Info Magnetic Resonance Imaging (MRI) Information for Grace Medical Otologic Implants ![]()